Mar 10, 2017  BROOKFIELD DV2T Viscometer kinematic viscosity dynamic viscosity water viscosity viscosity units viscosity meter viscosity measurement unit of. View and Download Brookfield DV-II+ operating instructions manual online. DIGITAL VISCOMETER. Measuring Instruments Brookfield DV2T Operating Instructions Manual. Touch screen viscometer (91 pages). The Brookfield Viscometer/Rheometer will produce an accurate and repeatable torque reading under any measurement circumstance. Leader in viscosity, texture analysis, and powder flow instrumentation! Brookfield has been considered the world standard in viscosity measurement and control for more than 80 years. Brookfield has a reputation of being the world leader in viscosity measurements for over 80 years providing a variety of laboratory viscometers to suit a wide variety of applications. ©2019 AMETEK.Inc. All rights reserved. By continuing to use this site, you agree to our Policy and Cookie policy. Brookfield Engineering Labs., Inc. Page 4 Manual No. M/03- 65-C0508 C0808 I. INTRODUCTION The Brookfield DV-II+Pro Viscometer measures fluid viscosity at given shear rates.Viscosity is a measure of a fluid’s resistance to flow.

  1. Rvdv2t Brookfield Viscometer
  2. Brookfield Dv2t Viscometer Manual Online
  3. Brookfield Viscometer Dv2t User Manual
  4. Brookfield Dv2t Viscometer Manual 2016
  5. Brookfield Dv2t Viscometer Manual Pdf
Advertising

Brookfield Engineering Labs., Inc.

Page 76

Manual No. M13-167-B0614

When a UL or DIN UL Adapter is used, the water bath is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the UL Tube (refer to the

UL Adapter instruction manual).

2) Attach the spindle (with extension link and coupling nut) onto the DV2T.
3) Attach the tube to the mounting channel.
4) Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket,

connect the inlet/outlets to the bath external circulating pump.

Brookfield Dv2t Viscometer Manual

5) Allow 30 minutes for the viscosity standard, sample chamber and spindle to

reach test temperature.

6) Measure the viscosity and record the Viscometer reading.

Note: The spindle must

rotate at least five (5) times before readings are taken.

Calibration Procedure using a Helipath Stand and T-Bar Spindles

When a Helipath Stand and T-Bar spindles are used:

1) Remove the T-bar spindle and select a standard LV (#1-3) or RV, HA, HB (#1-6)

spindle. Follow the procedures for LV (#1-3) and RV, HA, HB (#1-6) Brookfield

spindles outlined above.

2) T-Bar spindles should not be used for verifying calibration of the DV2T

EXCITING 3D ACTION 3D graphics and simple controls make for free-flowing and fast-paced hack-and-slash action. EPIC SPECIAL MOVES Unleash each Bleach character's unique special moves to carve your way to victory. Special moves are fully voiced by the original Japanese anime voice actors. THREE'S COMPANY Combine your favorite characters into teams of three. It’s a 3D action game with an addicting music and awesome fighting in different quests.Bleach Brave Souls PC Version is downloadable for Windows 10,7,8,xp and Laptop.Download Bleach Brave Souls on PC free with XePlayer Android Emulator and start playing now! How to Play Bleach Brave Souls on PC-Download and Install XePlayer Android Emulator. Bleach game download pc. Bleach games free download - Bleach It, Bleach Edition Manga Quiz - Ichigo Edition Japanese Game Free, Wikia: Bleach, and many more programs. Aug 20, 2019  Download Bleach Brave Souls 9.1.1. Official video game for Bleach the anime. Bleach Brave Souls is a 3D action game where players control some of the most important characters from the well-known anime Bleach. Ichigo, Byakuya, Renji, and Kenpachi among many others join forces for the same cause: fight.

Viscometer.

Calibration Procedure for Spiral Adapter

Rvdv2t Brookfield Viscometer

Brookfield recommends a two step check. First, verify the calibration of the viscometer

using the standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or

cone/plate spindles) as detailed in this appendix. Second, verify the calibration of the

viscometer using the Spiral Adapter. The use of an accessory device may increase the

accuracy of measurement associated with the DV2T.

1) Place the viscosity standard fluid (in the proper container) into the water bath

(refer to the Spiral Adapter instruction manual).

2) Attach the spindle to the viscometer. Attach chamber (SA-1Y) and clamp to the

viscometer.

3) Lower the DV2T into measurement position. Operate the viscometer at 50 or

60 RPM until the chamber is fully flooded.

4) The viscosity standard fluid, together with the spindle, should be immersed in

the bath for a minimum of 1 hour, stirring the fluid periodically (operate at 50 or

60 RPM periodically), prior to taking measurements.

5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate

thermometer.

6) If the fluid is at test temperature (

+

/- 0.1

°

C of the specified temperature, normally

25

°

C), measure the viscosity and record the viscometer reading.

Note: The spindle

must rotate at least five (5) times for one minute, whichever is greater before readings are

taken.

7) The viscosity reading should equal the cP value on the viscosity fluid standard

to within the combined accuracies of the viscometer and the standard (as

discussed in the section entitled, Interpretation of Calibration Test Results).

Advertising
Popular Brands
Popular manuals
Advertising

Brookfield Engineering Labs., Inc.

Page 5

Manual No. M13-167-B0614

I. INTRODUCTION

The Brookfield DV-II Viscometer series has been the leading industrial viscometer since it

was first introduced in 1985. Brookfield has continued to develop and improve the DV-II to

provide the best value in the market for both quality control and research customers. The

Brookfield DV2T Viscometer continues in this tradition of innovation, quality and value. The

incorporation of a full color graphical touch screen display has allowed for a new and improved

user interface which preserves the single speed data collection methods while making all of

the advanced features of the DV2T readily available.
The Brookfield DV2T Viscometer measures fluid viscosity at given shear rates. Viscosity is

a measure of a fluid’s resistance to flow. You will find a detailed description of the science of

viscosity in the Brookfield publication “More Solutions to Sticky Problems”, a copy of which

was included with your DV2T.
This manual covers the DV2T Viscometer and the DV2T Extra Viscometer. The DV2T Extra is

a special package that includes several optional features to maximize the utility of the DV2T.

The DV2T Extra includes: Ball Bearing Suspension (not included in LV versions), EZ-Lock

spindle connection, Model Q Lab Stand, and RheocalcT Software. These differences are

detailed in the Components section of the manual. All references to the viscometer in this

manual will be made as DV2T. All of the optional features of the DV2T Extra may be added

individually to the DV2T Viscometer.

• The DV2T offers exceptional versatility in modes of control, allowing choice of traditional

standalone operation and automatic operation through programs downloaded from

the PC or with complete control by PC using Brookfield RheocalcT Software.

• The DV2T can be used as a traditional Brookfield viscometer for collection of single

speed viscosity data through the easy to use touch screen; just select the spindle and

speed and read the value from the display. [See Section II: Getting Started]

• The Brookfield PG Flash Software can be used to program the DV2T to control all

aspects of the test and data collection without the need for the operator to monitor

the instrument; just start the program and return to the printed test data (printer is

optional). [See Section V: PG Flash Software]

• The Brookfield RheocalcT Software will perform all control and data collection functions

of the DV2T from the PC while also providing a platform for advanced data collection
and analysis.

In any of these modes of control, the DV2T will provide the best in viscosity measurement

and control.

The principal of operation of the DV2T is to drive a spindle (which is immersed in the test fluid)

through a calibrated spring. The viscous drag of the fluid against the spindle is measured by the

spring deflection. Spring deflection is measured with a rotary transducer. The measurement

range of a DV2T (in centipoise or milliPascal•seconds) is determined by the rotational speed

of the spindle, the size and shape of the spindle, the container the spindle is rotating in, and

the full scale torque of the calibrated spring.
There are four basic spring torque series offered by Brookfield:

Spring Torque

Model

dyne

•cm

milliNewton

•m

DV2TLV

673.7 0.0673

DV2TRV

7,187.0

0.7187

DV2THA

14,374.0

1.4374

DV2THB

57,496.0

Brookfield Dv2t Viscometer Manual Online

5.7496

The higher the torque calibration, the higher the measurement range. The measurement

range for each torque calibration may be found in Appendix B.

Brookfield Viscometer Dv2t User Manual

Advertising

Brookfield Dv2t Viscometer Manual 2016

Popular Brands

Brookfield Dv2t Viscometer Manual Pdf